Summary
How to run an 85°C/85%RH biased aging test: sample preparation, conditions, recovery, measurement and common mistakes.
What 85/85 means
85/85 testing exposes LEDs to 85°C and 85% relative humidity while electrically biased. It is used to evaluate moisture-related reliability.
Before testing
- Bake/dry the samples as specified
- Record initial luminous flux, color coordinates, VF and IR
- Use moisture-resistant fixtures to avoid short circuits
- Verify chamber temperature/humidity calibration
Test conditions
| Item | Setting |
|---|---|
| Temperature | 85°C |
| Humidity | 85% RH |
| Bias | Rated current or customer-specified |
| Duration | 168 / 500 / 1000 h |
| Recovery | About 2 h at room temperature before measurement |
During testing
- Do not apply bias until temperature and humidity are stable
- Watch for condensation, which can cause leakage or shorts
- Log chamber temperature and humidity
- Avoid frequent door opening
After testing
- Recover to room temperature before measurement
- Measure flux, color, VF and IR
- Inspect for yellowing, corrosion and dead LEDs
- Judge against the agreed criteria
Common mistakes
- Loading samples without drying
- Applying bias during ramp-up and causing condensation
- Measuring immediately after removal
- Checking brightness only, not IR or color shift
- No initial data