Summary
How to run a lumen decay test: fixed current, temperature and measurement conditions, data recording and lumen maintenance criteria.
What lumen decay testing measures
Lumen maintenance is the ratio of luminous flux at a given time to the initial flux. It is used to judge light output degradation and to support lifetime estimates.
Test conditions
| Item | Requirement |
|---|---|
| Current | Rated current, constant current |
| Temperature | 25°C / 55°C / 85°C |
| Time points | 0 / 24 / 48 / 168 / 500 / 1000 h |
| Measurement | Same current, temperature and warm-up time before and after aging |
| Recovery | Return to room temperature before measurement |
Data to record
- Luminous flux and maintenance ratio
- VF and IR
- Color coordinates / CCT
- Test temperature, current, equipment ID and operator
Typical criteria
| Parameter | Common criterion |
|---|---|
| Lumen maintenance | ≥95% at 500 h, ≥90% at 1000 h (per customer spec) |
| Color shift | Δu′v′ ≤ 0.005 |
| VF change | ≤ ±5% |
| IR | Within datasheet limit |
Criteria should be agreed before testing, not after.
Common mistakes
- Different measurement conditions before and after aging
- No temperature control
- Ordinary power supply instead of constant current
- Measuring immediately after aging
- Testing only one LED